Journal Search Engine
Search Advanced Search Adode Reader(link)
Download PDF Export Citaion korean bibliography PMC previewer
ISSN : 1225-0112(Print)
ISSN : 2288-4505(Online)
Applied Chemistry for Engineering Vol.9 No.6 pp.857-863
DOI :

RESEARCH PAPERS : A New Method for Determination of Enzyme Reaction and Activity of Lysozyme with UV - Spectrophotometer

Woon Soo Kim , Yong Wook Kim , Woo Sik Kim

Abstract

A simple and new experimental method for determination of lysozyme-M. lysodeikticus cell lysis reaction and lysozyme activity was suggested using Beer's law. The UV transmittance of the solution changed with the concentration of M. lysodeikticus and the relationship between the UV transmittance and M. lysodeikticus cell concentration followed Beer's Law. In addition, it was experimentally proven that the UV transmittance of the solution was not influenced by the lysozyme concentration and product of the lysis reaction. During the lysozyme-M. lysodeikticus cell lysis reaction, thus, M. lysodeikticus cell concentration in the solution could be measured in-situ by UV-spectrophotometer. By using these experimental data, kinetic Parameters of the Michaelis-Menten equation for the lysozyme-M. lysodeikticus cell 1ysis reaction was simply determined The maximum reaction rate constant (k3) and Michaelis-Menten constants were 0.1734 sec-1and 9.83 X 10-6M respectively. The activity of the lysozyme could also be obtained with this experiment because the lysis reaction rate of the 1ysozyme depended on its activity.

연구논문 : UV - 분광광도계를 이용한 새로운 Lysozyme 의 효소반응 및 활성측정법 연구

김운수,김용욱,김우식

초록

Beer의 법칙을 이용하여 lysozyme-M. lysodeikticus cell의 용해반응속도 및 lysozyme의 활성을 측정할 수 있는 간단하고 새로운 실험방법을 제시하였다. M.lysodeikticus cell의 농도에 따라 용액의 UV 투과도를 측정하였으며 이것은 Beer의 법칙에 따르는 것으로 나타났다. 또한 용액 내의 lysozyme 농도 및 용해반응에 의해 생성되는 생성물은 UV 투과도에 전혀 영향을 미치지 않는 것으로 실험적으로 증명되었다. 따라서 lysozyme의 용해반응이 진행되는 동안 용액내에 존재하는 M. lysodeikticus cell의 농도는 UV-분광광도계를 이용하여 in-situ로 측정될 수 있었다. 이렇게 얻어진 M. lysodeikticus cell의 농도변화를 Michaelis-Menten식을 이용하여 lysozyme-M. lysodeikticus cell의 용해 반응속도 상수들을 결정하였다. 최대반응속도상수 k3은 약 0.1734 sec-1로 나타났으며 Michaelis-Menten 상수는 약 9.83X10-6M으로 나타났다. Lysozyme의 용해반응속도가 활성에 의존하므로 이와 같은 측정을 통해 lysozyme의 활성도도 측정할 수 있었다.

Figure

Table